It is often not possible for a physician to administer medications that are effective at the lowest dosages with a customary needle. With syringe pumps, these highly effective drugs can, however, be administered through the vein for longer periods of time. The smallest amounts can be exactly dosed this way.
Fresenius Kabi’s infusion devices are also used to introduce infusion solutions. Syringe pumps precisely administer small quantities (0.1 to 200 milliliters per hour) while volumetric infusion pumps are used to deliver larger quantities (1 to 1,500 milliliters per hour). Fresenius Kabi is among the leading manufacturers of pumps for infusion and clinical nutrition in Europe.
Infusia SP 7
Flow rate range
50 ml syringe: From 0.1 to 1800.0 ml
30 ml syringe: From 0.1 to 1200.0 ml/h
20ml syringe: From 0.1 to 800.0 ml/h
10ml syringe: From 0.1 to 400.0 ml/h
5 ml syringe: From 0.1 to 200.0 ml/h
Increment
0.1 ml/h for all syringe types
Syringe capacities
5 ml, 10 ml, 20 ml (30 ml), 50 ml (60 ml)
Accuracy
±1% device accuracy (without disposable variability), ±3% total accuracy
Volume to be infused
From 0.0 to 1999.9 ml, 0.1 ml increment
Infusion time
From 1.0 min to 1999 min, 1 min increment
Bolus or Purge rate
50 ml syringe: From 0.1 to 1800.0 ml/h
30 ml syringe: From 0.1 to 1200.0 ml/h
20ml syringe: From 0.1 to 800.0 ml/h
10ml syringe: From 0.1 to 400.0 ml/h
5 ml syringe: From 0.1 to 200.0 ml/h
Increment
0.1 ml/h for all syringe types
Bolus volume
From 1.0 to 100.0 ml, 0.1 ml increment
Infusia VP 7
Flow rate range
1.0 ml/h - 1200 ml/h
1.0 - 99.9 ml/h, 0.1 ml/h increment
100 - 1200 ml/h, 1 ml/h increment
1 - 400 drops/min, 1 drop/min increment
Flow rate accuracy
± 5 %
Volume to infuse
0 - 9999 ml, 1 ml increment
Infusion mode:
Rate setting mode - ml/h, drops/min
Time setting mode - minutes, ml/h, drops/min
Volumetric setting mode - ml/h
Time of infusion
1 - 1999 min, 1 min increment minimum
Priming
Manual or by using the purge/bolus function of the pump
Bolus rate
1 - 1200 ml/h
0.1 - 99.9 ml/h, 0.1 ml/h increment
100 - 1200 ml/h, 1 ml/h increment
Purge rate
600 ml/h
Free flow protection
Anti-free flow clamp when door is closed
End of infusion:
Automatic KVO at the end of infusion: 1 ml/h if < 10 ml/h infusion rate, otherwise, 3 ml/h
Data log event
5,000 events